1. Verify that all power cables were properly connected.
2. Check that the main switch located on the left hand side of the analyzer is at “ON” position.
3. Check that the main fuses are not burnt.
4. Pull out the plug of power supply cable from its receptacle on the analyzer. Open up the fuse cover and pull if defective fuses are verified, replace it when necessary.
5. Check that the circuit breaker of the power supply system to which the analyzer is connected, if it is not damaged.
1. Check the following factors which affect results (Reagent preparation, Preparation of QC Sample, Preperation of Calibrator).
2. Check if all carterias are met in the following factors (on-board stability, expiry date, calibration status).
3. Check latest calibration factors if still within the given range, calibrate if necessary.
1. Check water supply quality.
2. Replace Disposable cuvettes and see if error perisist.
3. Check lamp. Lifespan of lamp is around 1000 hours. Replace if needed.
1. Check for obstructions detected by the sensor vertically.
2. If there are no visual obstructions observed check the sensor if the metal plate of the probe asssembly is in between the sensor.
3. Reboot the entire system in order to initialize the hardwares and sensors.
1. Make sure that the correct port has been selected in Service Check (under PC Communication).
2. Check continuity status of RS232 connector, might be disconnected or faulty, replace if found defective.
3. Reboot the entire system and verify the operational status.
1. Check the quality of DI water being used.
2. Check for clogs in the tubes.
3. Manually soak tubes for clogs.
4. Verify that the reagent carousel was properly closed to maintain refrigeration temperature. Water drains if the temperature is above the desired value.
1. Verify that the sample or reagent is loaded correctly and that the appropriate detection method is selected.
2. Check for any blockages or leaks in the fluidic system.
3. Make sure that the probe tip is clean and free from serum residue in order to detect the sample and the reagent volumes.
1. Check for loose or damaged parts and tighten or replace them as necessary.
1. Check if the instrument’s memory is full, back-up files and delete unnecessary data.
2. Verify if other applications or processes are running concurrently, as they may be affecting the instrument’s performance.
1. Ensure that the barcode is intact and properly positioned for scanning.
2. Clean the barcode scanner if necessary. If the error persists, manually enter the sample information.
3. Check barcodes from sample tubes, might be torn and could no longer be scanned. Replace when needed or manually input the details.
1. Check for air bubbles or contaminants in the sample. Mix the sample thoroughly before testing.
2. Ensure that the sample and reagents are at the correct temperature and that the appropriate test protocol is followed (on-board stability, expity date).
1. Check the sample probe for any clogs or blockages.
2. Clean the probe with alcohol or cleaning solution if necessary.
3. Ensure that the sample volume is adequate with the number of test required.
1. Verify that all power cables were properly connected.
2. Check that the main switch located on the left hand side of the analyzer is at “ON” position.
3. Check that the main fuses are not burnt.
4. Pull out the plug of power supply cable from its receptacle on the analyzer. Open up the fuse cover and pull if defective fuses are verified, replace it when necessary.
5. Check that the circuit breaker of the power supply system to which the analyzer is connected, if it is not damaged.
1.Check the following factors which affect results (Reagent preparation, Preparation of QC Sample, Preperation of Calibrator).
2. Check if all carterias are met in the following factors (on-board stability, expiry date, calibration status).
3. Check latest calibration factors if still within the given range, calibrate if necessary.
1. Check water supply quality.
2. Replace Disposable cuvettes and see if error perisist.
3. Check lamp. Lifespan of lamp is around 1000 hours. Replace if needed.
1. Check for obstructions detected by the sensor vertically.
2. If there are no visual obstructions observed check the sensor if the metal plate of the probe asssembly is in between the sensor.
3. Reboot the entire system in order to initialize the hardwares and sensors.
1. Make sure that the correct port has been selected in Service Check (under PC Communication).
2. Check continuity status of RS232 connector, might be disconnected or faulty, replace if found defective.
3. Reboot the entire system and verify the operational status.
1. Check the quality of DI water being used.
2. Check for clogs in the tubes.
3. Manually soak tubes for clogs.
4. Verify that the reagent carousel was properly closed to maintain refrigeration temperature. Water drains if the temperature is above the desired value.
1. Verify that the sample or reagent is loaded correctly and that the appropriate detection method is selected.
2. Check for any blockages or leaks in the fluidic system.
3. Make sure that the probe tip is clean and free from serum residue in order to detect the sample and the reagent volumes.
1. Check for loose or damaged parts and tighten or replace them as necessary.
1. Check if the instrument’s memory is full, back-up files and delete unnecessary data.
2. Verify if other applications or processes are running concurrently, as they may be affecting the instrument’s performance.
1. Ensure that the barcode is intact and properly positioned for scanning.
2. Clean the barcode scanner if necessary. If the error persists, manually enter the sample information.
3. Check barcodes from sample tubes, might be torn and could no longer be scanned. Replace when needed or manually input the details.
1. Check for air bubbles or contaminants in the sample. Mix the sample thoroughly before testing.
2. Ensure that the sample and reagents are at the correct temperature and that the appropriate test protocol is followed (on-board stability, expity date).
1. Check the sample probe for any clogs or blockages.
2. Clean the probe with alcohol or cleaning solution if necessary.
3. Ensure that the sample volume is adequate with the number of test required.
1. If the COM cable is loose, fasten the cable.
2. For improper COM, select a proper COM.
3. If problem persists, please contact JARCET.
1. Replace the fuse (T0.8A delay fuse).
2. If problem persists, please contact JARCET.
1. Please contact JARCET.
1. Reinstall pump if the pump is not installed well.
2. Replace pump if the pump has deteriorated or cracked.
3. Reset up parameter for the pump for improper pump parameter set.
4. If failed online, press 1 to check up.
5. If promlem persists, please contact JARCET.
1. Check up the leakage and reconnect the tubes.
2. Replace the pump.
1. Use the onboard cable if the COM cable is too long.
2. Ensure that the cable is fastened correctly.
3. If there is strong electromagnetic interference nearby, remove the high power equipment.
4. If problem persists, please contact JARCET.
1. If the lamp is broken, please contact JARCET.
2. Drain off the liquid int he cell and repipette liquid to adjust in case of air bubbles or insufficient water in the cell. If unsuccessful, reset up parameter for the pump.
3. Rinse the cell with detergent or absolute alcohol several times.
4. If problem persists, please contact JARCET.
1. Reset the parameter for the pump if there are air bubbles in the cell.
2. The silica get pipe may have deteriorated or cracked, resulting in inaccurate pipetted volume. If this is the case, replace silica get pipe for the pump.
3. Readjust the initial A/D in case of ageing with the lightsource lamp.
4. For improper item test parameter set, reset item parameter according to the reagent instructions.
5. In the absence of earth wire with the power supply, ground the instrument with the onboard ground pole.
6. If problem persists, please contact JARCET.