FAQs - Coagulation Products
Coag 2D Coagulometer
OUT OF RANGE CONTROLS
Check if it is caused by instrument reagent or method failure.
INACCURACY OR DEVIATION BETWEEN CHANNELS
Light paths may be dirty or contaminated. Check reproducibility with parallel measurements.
ABNORMAL MEASUREMENT START OR STOP
Check with standard (PT) assay and normal sample.
FROZEN SOFTWARE
Switch off and restart.
TEMPERATURE REGULATION FAILURE
Switch off.
Call service support.
DIFFERENCE ERROR / OUT OF RANGE ERROR / CURVE ERROR
Repeat test.
EXTERNAL LIGHT ERROR
Restart analyzer.
CALIBRATION DATA ERROR
Input calibration data.
INCUBATION IS TOO LONG
Repeat test.
OVERHEATING INCUBATION
Restart analyzer.
EXPIRED LOT
Use another lot.
OUT OF QC
Re-calibrate.
Coag 4D Coagulometer
OUT OF RANGE CONTROLS
Check if it is caused by instrument reagent or method failure.
INACCURACY OR DEVIATION BETWEEN CHANNELS
Light paths may be dirty or contaminated. Check reproducibility with parallel measurements.
ABNORMAL MEASUREMENT START OR STOP
Check with standard (PT) assay and normal sample.
FROZEN SOFTWARE
Switch off and restart.
TEMPERATURE REGULATION FAILURE
Switch off.
Call service support.
DIFFERENCE ERROR / OUT OF RANGE ERROR / CURVE ERROR
Repeat test.
EXTERNAL LIGHT ERROR
Restart analyzer.
CALIBRATION DATA ERROR
Input calibration data.
INCUBATION IS TOO LONG
Repeat test.
OVERHEATING INCUBATION
Restart analyzer.
EXPIRED LOT
Use another lot.
OUT OF QC
Re-calibrate.
Coag M Coagulometer
AN ERROR HAS OCCURED WHILE SENDING LOG FILES
Check the network connection and the internet access.
BACKUP ERROR
Check the backup device and try again.
BARCODE READER DOESN'T REPLY
Make sure that the barcode reader light is on. Do not stare into it directly. Try to read another barcode.
CLEANER NOT DETECTED
The reagent or sample arm does not sense Cleaner at the washing station.
Check the Cleaner at washing station during washing process. Clean the washing station by Dia-Clean-Sys.
CLEANER TANK IS EMPTY OR WASTE IS FULL DURING MEASUREMENT RUN
Check the Cleaner tank and Waste tank.
Replace the Cleaner tank and Waste tank.
CLEANER TANK IS EMPTY BEFORE READY STATE
Replace the Cleaner tank and Waste tank.
CLEANER TANK IS EMPTY DURING PRIME OR WASH
The Cleaner tank is empty.
Check the level of Cleaner liquid. Replace the Cleaner tank.
CLEANER TANK HAS NOT BEEN CHANGED
You didn’t screw down and pull out the transfer tube signed with green colour.
During changing the Cleaner, screw down, then pull out the transfer tube signed with green colour.
INCOMPLETE OR EMPTY REAGENT
One or more reagents are missing required for the measurements.
Please check the list of reagents. Add the missing reagents and run the measurement.
FAILED CUVETTE DROP
The Cuvette arm position is wrong or the tweezers are dirty.
Check visually the position of Cuvette arm, and clean the tweezers.
EMPTY CUVETTE HOPPER
The cuvettes run out, or the volume indicated in the program is zero.
Check the volume of cuvettes in the Cuvette Hopper. Add cuvettes and type the correct number.
CUVETTE HOPPER ROTOR JAM
The Cuvette incubator rotor position is wrong, the rotor is not moving properly.
Check visually the position of Cuvette incubator rotor. Clean the rotor and start initialisation.
CUVETTE IS MISSING IN INCUBATOR ROTOR
The Cuvettes in incubator is stacked, or the cuvette is not detected properly.
Check visually the position of Cuvettes in incubator. Clean the incubator and start initialisation.
CUVETTE IS NOT DETECTED AFTER SHAKING
There was no cuvette taken back to the measuring position. Check visually the position of Cuvette arm, and clean the tweezers.
LOW LEVEL
The Cuvette incubator rotor position is bad or the cuvette is not detected properly.
Check visually the position of Cuvette incubator rotor. Clean the rotor and start initialisation.
CUVETTE IS NOT DETECTED ON MEASPOS
There was no cuvette taken to the measuring position.
Check visually the position of Cuvette arm, and clean the tweezers.
CUVETTE TRASH IS FULL
The trash is full or the volume indicated in the program is 1000.
Empty the trash.
DIACLEAN-SYS LEVEL IS TOO LOW
The needles of reagent and sample arms sense low liquid level.
Check the level of Dia-Clean-Sys. Fill up the vials.
DISABLED USER
The user name you tried to enter is not valid.
Check the list of user names. Try another user.
EMERGENCY STOP EXECUTED
The instrument stopped after emergency stop and the measurements have lost.
Restart the program and repeat the lost measurements.
EXPIRED QC PERIOD / OUT OF RANGE QC VALUES
The QC period of reagent has expired or the QC value out of range.
Run new control.
INCORRECT ACCESS LEVEL
The level of the user you applied to the Coag M is not enough for the process.
Check the user. Choose another user.
INCORRECT CALIBRATOR NAME
The word “Barcode” is not allowed for calibrator name.
Check the name of the calibrator. Type again.
INCORRECT DATE
The format of the date you typed in was incorrect.
Please check the format and repeat it.
INCORRECT SHUTDOWN OCCURRED
The incorrect shutdown may result wrong or inconsistent database.
Turn off and on the Coag M and check the data in the program.
INCORRECT TEST SETUP
One or more field is not correct in the Test Setup.
Check the field and type again.
INIPARAMS HAVE CHANGED
One or more Ini parameters were changed.
Exit Coag M and start again.
INITIALIZATION ERROR
There was an error during the initialization process.
Check the part what was not initialized. Start initialization again. Have the service maintain the instrument.
INSERTED REAGENTS DON'T MATCH TO LAST STATE
The Coag M checked the position of reagents during initialization. The position of reagents do not match to the list exists in the program.
Check the reagents and type into window appear step by step.
INSTRUMENT IS NOT IN READY STATE
You tried to start a process what was denied.
Wait for the Ready state, and then try again.
LEVEL DETECTION ERROR
There is a level detection error.
Check the liquid in the vial at given place. Try again.
LOT NUMBERS MISMATCH
The scanned data not belong to the reagent/control/calibrator.
Check the data sheet. Try again with correct data sheet.
LOW LEVEL
The liquid level of the reagent is lower than the predetermined percent of the max volume.
Prepare new bottle of reagents and add to the reagent holder.
MACHINE IS UNINITIALIZED
The instrument is not initialized after certain changes.
Run the initialization.
MAXIMUM CAPACITY OF THE HOPPER IS CUVETTES
The cuvette number exceeds the 500.
Do not add more than 500 cuvettes.
MEASURED VOLUME IS DIFFERENT THAN GIVEN VOLUME
The volume of the reagents listed in the reagent table, differ from the volume calculated from level sensing. Press Yes to accept the calculated volume.
MISSING CALIBRATION
The calibration of the … reagent is missing.
Please add calibration data or arrange auto calibration before start to measure.
NO CONTROL INSTRUCTION IN THE TEST SETUP
There is no control data in the Test Setup.
Add control data to the Test Setup.
NO CUVETTE AT SAMPLE POSITION
There must be cuvette at sample position in rotor during sampling.
Please check the rotor of incubator and the cuvette feeder.
NOT ENOUGH SOLUTIONS
There are no enough reagents for the planned measurements.
Check the reagents. Add new reagents for the measurement.
NOT ENOUGH, MISSING, OR EXPIRED CONTROL
There is no enough control for the planned QC measurements.
Check the controls. Add controls for the QC measurement.
OPTICAL RACK DETECTION FAILED
The rack sensors of rack changer failed to detect the rack properly.
Have the service maintain the instrument.
QC VALUE IS OUT OF THE ALLOWED RANGE
The measured QC value is out of range.
Check the preparation and measurement of QC and try again. Have the service maintain the instrument.
RACK HOME SENSOR ERROR
The mechanic of rack loader has failed.
Start initialization process. Clean the surface or rack and rack loader.
RACK MOTOR HOME ERROR
The movement of rack motor is failed.
Start initialization process. Clean the surface or rack and rack loader.
RACK MOVING ERROR
The movement of rack was failed.
Start initialization process. Clean the surface or rack and rack loader.
RACK ID MUST BE SET
The Rack ID is empty.
You should add the Rack ID to the window.
SAMPLE ID READING ERROR AT SAMPLE POSITION
The sample barcode was not read properly.
Pick up the sample tube and try to scan it manually.
SAMPLE ID MUST BE SET
The measuring process needs the sample ID.
Add sample ID to the input window.
SAMPLING HAS STOPPED
Sampling has stopped because of some problems (e.g. Cleaner runs out) or manually stopped.
Wait for finishing the running measurements.
SMP ID ALREADY EXISTS ON A DIFFERENT POSITION
The system has recognised that a sample ID occur more than once.
Please always use unique sample ID for each samples.
SMP ID MISMATCH
The sample ID differ from the correct one.
Please check the rack loader.
SMP ID OCCURS MORE THAN ONCE
The sample ID must be unique.
Choose unique sample ID.
SMP ID ALREADY EXISTS IN THE WORKLIST
The sample already measured and results are in the Worklist.
You have to accept or delete the results in worklist, before you start the same samples.
STALL GUARD
One of the motors of the instrument has stacked.
Start initialisation.
TEMPERATURE ERROR
The temperature of units must be in certain range.
Please wait until all temperatures reach desired value.
CLEANER LOT NUMBER CAN NO LONGER BE USED
This Cleaner already was in the instrument. Cannot use more.
Take another Cleaner.
TRASH IS OPEN
The trash of cuvettes was opened during measurement.
The trash should be closed during operation.
TRAY IS FULL
The left side (input/output) of the tray is full.
Remove one or more racks from left side.
UNKNOWN RACK ID
Scanned Rack ID does not exist in the table.
Use another rack.
UNKNOWN REAGENT
The reagent you have tried to add by barcode was not allowed.
Use another reagent or call service.
WASTE TANK IS FULL
The waste tank is full.
Change the Cleaner and the waste according the instruction.
WASTE TANK HAS NOT CHANGED
You should change waste tank according to instructions.
Coagulyzer 1 Coagulation Analyzer
EXCEEDED MAXIMUM MEASURING TIME (BREAK TIMEOUT)
No clotting found was found. Perform optical test for clots.
Repeat test.
MEASUREMENT PREPARATION IS TOO TURBID (BREAK DARK)
Dilute plasma or mix reagent.
EXCEEDED MEASURING RANGE - TOO HIGH (BREAK TOP LIM)
Repeat test.
EXCEEDED MEASURING RANGE - TOO LOW (BREAK BOT LIM)
Repeat test.
MIXER MOTOR ERROR OCCURRED (BREAK MOTOR)
Contact technical service.
LOUD NOISE AFTER SAMPLE ADJUSTMENT (BREAK NOISE)
Check for air bubbles or other particles.
MEASURED CURVE DRIFTED AFTER ADDED REAGENT (BREAK DRIFT)
Check sample for air bubbles.
MEASUREMENT CANCELLED WITH RESET
This can only be caused by the user.
MEASURING BREAK DUE TO MEASURING VALUE JUMP / NO CLOTTING (BREAK JUMP)
Repeat test.
If necessary, contact technical service.
LIGHT VALUE IS TOO DARK DURING THE ADJUSTMENT PHASE (BREAK READJUST)
Repeat test.
Dilute plasma if necessary.
CONVERTED VALUE IS LOWER THAN PARAMETERIZED MIN VALUE
Check analytical steps and the conversion parameter.
CONVERTED VALUE IS HIGHER THAN PARAMETERIZED MAX VALUE
Check analytical steps and the conversion parameter.
LOGARITHM CONVERSION COMP. DIVIATION FROM NEG. VALUE COMP. OVERFLOW
Check conversion parameter.
If necessary, contact technical service.
SYSTEM FAILURE
Contact technical service.
Coagulyzer 2 Coagulation Analyzer
EXCEEDED MAXIMUM MEASURING TIME (BREAK TIMEOUT)
No clotting found was found. Perform optical test for clots.
Repeat test.
MEASUREMENT PREPARATION IS TOO TURBID (BREAK DARK)
Dilute plasma or mix reagent.
EXCEEDED MEASURING RANGE - TOO HIGH (BREAK TOP LIM)
Repeat test.
EXCEEDED MEASURING RANGE - TOO LOW (BREAK BOT LIM)
Repeat test.
MIXER MOTOR ERROR OCCURRED (BREAK MOTOR)
Contact technical service.
LOUD NOISE AFTER SAMPLE ADJUSTMENT (BREAK NOISE)
Check for air bubbles or other particles.
MEASURED CURVE DRIFTED AFTER ADDED REAGENT (BREAK DRIFT)
Check sample for air bubbles.
MEASUREMENT CANCELLED WITH RESET
This can only be caused by the user.
MEASURING BREAK DUE TO MEASURING VALUE JUMP / NO CLOTTING (BREAK JUMP)
Repeat test.
If necessary, contact technical service.
LIGHT VALUE IS TOO DARK DURING THE ADJUSTMENT PHASE (BREAK READJUST)
Repeat test.
Dilute plasma if necessary.
CONVERTED VALUE IS LOWER THAN PARAMETERIZED MIN VALUE
Check analytical steps and the conversion parameter.
CONVERTED VALUE IS HIGHER THAN PARAMETERIZED MAX VALUE
Check analytical steps and the conversion parameter.
LOGARITHM CONVERSION COMP. DIVIATION FROM NEG. VALUE COMP. OVERFLOW
Check conversion parameter.
If necessary, contact technical service.
SYSTEM FAILURE
Contact technical service.
Coapresta 2000 Coagulation Analyzer
CUVETTE NOT SET
Open cuvette cover/analyzer is on state.
Use cotton buds or soft gauze to clean two sensors nearby.
Observe cuvette hopper if rotates during cleaning.
Return cuvette cover.
Perform start up.
Proceed running of samples once analyzer is shown on ready state mode.
CUVETTE JAM
Open cuvette cover.
Check and remove cuvette that block the path going to reading chamber.
Return cuvette cover.
Perform start up and washing.
Proceed running of samples once analyzer is shown on ready state mode.
WASTE ALARM
Remove and check waste box if full or empty.
Clean surface internal and external of waste box.
Clean inner waste sensors with soft tissue or cloth.
Return waste box and closed the door.
Proceed running of samples once analyzer is shown on ready state mode.
ANALYZER NO RESPONSE
Switch off the analyzer.
Shutdown computer from Main menu.
Switch on analyzer and wait for a beep.
Turn on CPU.
Proceed running of samples once analyzer is shown on ready state mode.