EMPTY DETERGENT / LYSE / DILUENT
2. Check reagents tubings and connections.
3. Go to maintenance and perform prime twice.
4. Perform blank test and observe tubings flow of reagents going to the analyzer.
5. Proceed running of samples.
CLOGGED WBC AND RBC
2. Click cauterize twice.
3. Click flush aperture twice.
4. Perform Blank test and observe results and alarm.
5. Proceed running of samples once no alarm and blank test results is within acceptable range.
TEMPERATURE OUT OF RANGE
2. Turn off the analyzer and turn on a few minutes.
3. Perform blank test and observe the anlyzer alarm and results.
4. Proceed running of samples once no alarm and blank test is passed.
HIGH BLANK VALUE
2. Go to maintenance.
3. Click rinse fluid system twice.
4. Click rinse sample cup twice.
5. Perform blank test and check results if its within acceptable range.
6. Proceed running of sample once all results is passed.
INITIALIZATION FAILED
2. Reseat LAN cable.
3. Turn on analyzer.
4. Check results of initialization if its all passed.
5. Perform blank test and check acceptable results.
6. Proceed running of samples.
NO DIFFERENTIAL COUNT
2. Click ‘Clean Sheath Channel’.
3. Click ‘Soaked WOC Cup’.
4. Perform blank test and observe results.
5. Proceed running of samples.
EMPTY DETERGENT / LYSE / DILUENT / SHEATH
2. Check reagents tubings and connections.
3. Go to maintenance and perform prime twice.
4. Perform blank test and observe tubings flow of reagents going to the analyzer.
5. Proceed running of samples.
WBC / RBC BUBBLES AND CLOG
2. Click ‘Cauterize Apperture’.
3. Click ‘Flush Apperture’.
4. Click ‘Soaked WBC/RBC Cup’.
5. Perform prime twice.
6. Perform blank test and check acceptable results.
7. Proceed running of samples once results is passsed.
ANALYZER DOES NOT TURN ON
2. Check if the AVR / UPS is turned ON.
3. Check if the power cable of the analyzer is connected in the AVR / UPS.
4. Check if the analyzer power switch located at the back of the analyzer is turned ON.
WBC CLOG
2. Go to “FUNC” then press “H.V CAUTERY”. When the procedure is done run a blank test. If the problem still persist. Go to the next step.
3. Prepare a 4ml probe cleaner in a plain test tube. Go to “FUNC” then submerge the sample probe in the plain test tube with probe cleaner then press Tubing Clean. After the “PROBE CLEANING” procedure press “PRIME” after reagent priming. Run a Blank test.
RBC CLOG
2. Go to “FUNC” then press “H.V CAUTERY”. When the procedure is done run a blank test. If the problem still persist. Go to the next step.
3. Prepare a 4ml probe cleaner in a plain test tube. Go to “FUNC” then submerge the sample probe in the plain test tube with probe cleaner then press Tubing Clean. After the “PROBE CLEANING” procedure press “PRIME” after reagent priming. Run a Blank test.
WBC BUBBLES
2. Perform reagent priming. Press “FUNC” the press “PRIME” to eliminate any air gap in the reagent tubings. Do this for 2 to 3 times.
3. Ask the user to run a blank test (pressing of run button without presenting any sample).
RBC BUBBLES
2. Perform reagent priming. Press “FUNC” the press “PRIME” to eliminate any air gap in the reagent tubings. Do this for 2 to 3 times.
3. Ask the user to run a blank test (pressing of run button without presenting any sample).
VACUUM LOW
2. Perform reagent priming. Press “FUNC” the press “PRIME”. Do this for 2 to 3 times.
3. Ask the user to run a blank test (pressing of run button without presenting any sample).
WASTE FULL
2. Empty the waste container. If it is almost full.
DILUENT EMPTY
2. Perform reagent priming. Press “FUNC” the press “PRIME”. Do this 2 times.
3. Ask the user to run a blank test (pressing of run button without presenting any sample).
DETERGENT EMPTY
2. Perform reagent priming. Press “FUNC” the press “PRIME”. Do this 2 times.
3. Ask the user to run a blank test (pressing of run button without presenting any sample).
LYSE EMPTY
2. Perform reagent priming. Press “FUNC” the press “PRIME”. Do this 2 times.
3. Ask the user to run a blank test (pressing of run button without presenting any sample).
ANALYZER DOES NOT TURN ON
WBC CLOG
RBC CLOG
WBC BUBBLES
RBC BUBBLES
VACUUM LOW
WASTE FULL
DILUENT EMPTY
DETERGENT EMPTY
LYSE EMPTY
SHUTTING DOWN ANALYZER
ANALYZER DOES NOT TURN ON
WBC CLOG
RBC CLOG
WBC BUBBLES
RBC BUBBLES
VACUUM LOW
WASTE FULL
DILUENT EMPTY
DETERGENT EMPTY
LYSE EMPTY
NO RESPONSE AFTER SELECTING SOFTWARE ITEM
WBC / RBC BUBBLES
CLOGGED WBC AND RBC
CANNOT ASPIRATE SAMPLE
CLOGGED WBC / RBC
POWER WON'T TURN ON
NO REAGENT ASPIRATION
HIGH WBC, RBC, AND PLT BACKGROUND RESULT
BARCODE NOT ACCEPTING REAGENT DURING REAGENT CHANGE
WBC CHAMBER CLOG
RBC CHAMBER CLOG
CYCLE BUSY
NO DILUENT
OUT OF RANGE REAGENT TEMPERATURE
SYSTEM BUSY
POWER WON'T TURN ON
NO REAGENT ASPIRATION
HIGH WBC, RBC, AND PLT BACKGROUND RESULT
BARCODE NOT ACCEPTING REAGENT DURING REAGENT CHANGE
WBC CHAMBER CLOG
RBC CHAMBER CLOG
CYCLE BUSY
NO DILUENT
OUT OF RANGE REAGENT TEMPERATURE
SYSTEM BUSY
INSTALLATION PROBLEM
A CONTROL HAS FAILED. AUTO STOP IS NOT ENABLED.
A CONTROL HAS FAILED. AUTO STOP TRIGGERED.
2. Contact technical service.
A MATCHING PATIENT SPECIMEN WAS FOUND. STUDIES MODE IS ENABLED.
A NULL OR EMPTY XML STRING WAS RECEIVED FOR PARSING.
A/D OR MRC REFERENCE VOLTAGE IS APPROACHING THE OPERATING LIMITS
A/D OR MRC REFERENCE VOLTAGE EXCEEDED THE OPERATING LIMITS
A/D RESOURCE BUSY
A/D RESOURCE INSUFFICIENT PAIRS
A/D RESOURCE NOT MONITORING
A/D RESOURCE NOT READY
A/D RESOURCE TIMEOUT
AMBIENT TEMPERATURE EXCEEDED THE OPERATING LIMITS
AMBIENT TEMPERATURE IS APPROACHING THE OPERATING LIMITS
ASPIRATION SYRINGE PUMP MOVE COULD NOT BE VERIFIED
ATTEMPT TO ACKNOWLEDGE RESULTS RECEIPT FAILED
AN INVALID PRIMARY ID WAS RECEIVED
AN INVALID SECONDARY ID WAS RECEIVED
AN XML PARSER WARNING HAS OCCURRED
AN XML PARSING ERROR OCCURRED
AN XML PARSING ERROR OCCURRED: INVALID ATTRIBUTE VALUE
AN XML PARSING ERROR: UNKNOWN START TAG IN OBJECT
AN XML PARSING WARNING: UNKNOWN END TAG IN OBJECT
APPLICATION EXITED WHILE BACKUP WAS IN PROGRESS
ASPIRATION ERROR
a. The volume was sufficient
b. It does not contain clots or fibrin.
c. It was collected and stored properly.