1. Check if the AVR / UPS is connected in the power outlet.
2. Check if the AVR / UPS is turned ON.
3. Check if the power cable of the analyzer is connected in the AVR / UPS.
4. Check if the analyzer power switch located at the back of the analyzer is turned ON
1. Click the ALARM to stop the alarm sound.
2. Ask the user to run a blank test (pressing of run button without presenting any sample). If the WBC clog is still pressent please do the next step.
3. Click “FUNC” then go to “MAINT”, in the maintenance screen select “CAUTERIZE APPERTURE”. Wait until it is finish then click FLUSH APPERTURE. Ask the user to run a blank test (pressing of run button without presenting any sample). If the problem still persist. Go to the next step.
4. Prepare 3ml probe cleaner into a clean plain test tube.
5. Click “FUNC” then go to “MAINT”, in the maintenance screen select “RINSE FLUIDICS”, wait for the confirm pop up to apear before clicking YES put the test tube with probe cleaner beneath the sample probe for the first time (insure that the sample probe touch the bottom of the test tube).
6. Wait for the confirm pop up to re apear, before clicking YES put the test tube with probe cleaner beneath the sample probe for the second time (insure that the sample probe touch the bottom of the test tube). Wait for the rinse fluidics procedure to finish.
7. Ask the user to run a blank test (pressing of run button without presenting any sample).
1. Click the ALARM to stop the alarm sound.
2. Ask the user to run a blank test (pressing of run button without presenting any sample). If the RBC clog is still pressent please do the next step.
3. Click “FUNC” then go to “MAINT”, in the maintenance screen select “CAUTERIZE APPERTURE”. Wait until it is finish then click FLUSH APPERTURE. Ask the user to run a blank test (pressing of run button without presenting any sample). If the problem still persist. Go to the next step.
4. Prepare 3ml probe cleaner into a clean plain test tube.
5. Click “FUNC” then go to “MAINT”, in the maintenance screen select “RINSE FLUIDICS”, wait for the confirm pop up to apear before clicking YES put the test tube with probe cleaner beneath the sample probe for the first time (insure that the sample probe touch the bottom of the test tube).
6. Wait for the confirm pop up to re apear, before clicking YES put the test tube with probe cleaner beneath the sample probe for the second time (insure that the sample probe touch the bottom of the test tube). Wait for the rinse fluidics procedure to finish.
7. Ask the user to run a blank test (pressing of run button without presenting any sample).
1. Check if there is remaing detergent reagent, replace if it is already depleted.
2. Perform reagent priming. Press “PRIME” button located in front of the analyzer. Do this for 2 to 3 times.This will eliminate air gaps in the hydraulic system
3. Ask the user to run a blank test (pressing of run button without presenting any sample).
1. Check if there is remaing detergent reagent, replace if it is already depleted.
2. Perform reagent priming. Press “PRIME” button located in front of the analyzer. Do this for 2 to 3 times.This will eliminate air gaps in the hydraulic system.
3. Ask the user to run a blank test (pressing of run button without presenting any sample).
1. Perform reagent priming. Press “PRIME” button located in front of the analyzer. Do this for 2 to 3 times. This will make the analyzer build up vacuum in the hydraulic system.
2. Ask the user to run a blank test (pressing of run button without presenting any sample).
1. Check the wasted level in the waste container.
2. Empty the waste container. If it is almost full.
1. Check if there is remaining Diluent in the container. Replace the diluent reagent If it is already depleted.
2. Click “FUNC” then “MAINT”. Select “PRIME DILUENT”. Do this 2 times.
3. Ask the user to run a blank test (pressing of run button without presenting any sample).
1. Check if there is remaining Detergent in the container. Replace the detergent reagent If it is already depleted.
2. Click “FUNC” then “MAINT”. Select “PRIME DETERGENT”. Do this 2 times.
3. Ask the user to run a blank test (pressing of run button without presenting any sample).
1. Check if there is remaining Lyse in the container. Replace the Lyse reagent If it is already depleted.
2. Click “FUNC” then “MAINT”. Select “PRIME LYSE”. Do this 2 times.
3. Ask the user to run a blank test (pressing of run button without presenting any sample).
1. Click EXIT icon located in the lower right most sid of the LCD.
2. Confirm pop up will appear. Click YES.
3. Confirm pop up will appear again this time it will ask you if you want to perform TUBING Clean. Prepare 3ml of probe cleaner into a plane test tube. Place the test tube under the aspiration probe and click YES. Wait for the sample probe to go up before removing the test tube.
4. Wait for the THANK YOU, NOW TURN OFF POWER message in the LCD. If this message appeared you can now turn OFF the power switch located at the back of the analyzer.
5. Don’t forget to turn OFF the UPS / AVR.