HbA1c Analyzer Support Questions
HPLCline 100 HbA1c Analyzer
COMMUNICATION TROUBLE
- Check whether the connection between the analyzer and the receiver is normal.
- The computer terminal and the analyzer IP address are consistent.
- Open the analyzer interface LIS transfer switch.
BARCODE SCANNING GUN COMMUNICATION FAILURE
- Check whether the barcode scanning gun is properly connected.
- Check if the connection is properly connected.
INVALID BARCODE
- Contact a JARCET service engineer for assistance.
NO HEMOLYTIC AGENT
- Check whether the hemolytic connection tube is well connected.
- Replace the new hemolytic agent.
NO ELUENT (A, B, C)
- Replace the solution.
- Reset liquid count.
EXISTING BUBBLES IN THE LINE CONNECTED REAGENTS
- Check the connection tube and reagent dosage, then perform to prime A/B/C solution to remove the bubbles.
RESULTS RANGED FROM 3% TO 18%
- Increase the sample size.
- Replace the whole blood sample or adopt the diluted blood model.
- Replace the column or filter.
- Reagent.
- Bubble removal.
ERROR CODE E1
- Mix it well and repeat the test. If the retest value is still abnormal, the guidance recommends testing with the method of diluting blood.
- Reposition the sample and retest it
ERROR CODE E2
- Mix it well and repeat the test. If the retest value is still abnormal, the guidance recommends testing with the method of diluting blood.
- After performing the bubble action, retest the sample.